Article

MIDSX: A Monte Carlo Interaction and Dosimetry Simulation of X-Rays

AUG 01, 2024
John Meneghini

Abstract: Computed tomography (CT) imaging, while essential for diagnostics, exposes patients to ionizing radiation. To accurately quantify radiation dosage, this study introduces MIDSX, a specialized open-source Monte Carlo (MC) photon transport code system for medical imaging. Unlike general-purpose particle transport MC systems, MIDSX is tailored for x-ray transport, offering more streamlined implementation. Results from validation simulations show MIDSX’s results for specific cases agree to within 0.39% of the mean of established reference data, demonstrating excellent agreement. However, discrepancies in body energy deposition measurements, reaching up to a 6.3% mean percent error, indicate areas for further refinement in the system.

https://doi.org/10.1063/10.0034186

This Content Appeared In
/
Issue
JURP 2024 Cover.png